Olympus Bronchovideoscope, Models BF-P190 & BF-XP190.

FDA device recall Z-2500-2023 · posted 2023-08-31 · Open, Classified

Recall details

Recalling firm
Olympus Corporation of the Americas
Reason for recall
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Action taken
Consignees were mailed an Urgent Medical Device Correction notification dated 7/27/2023 via UPS. The notification asks that consignees inspect their inventory for recalled devices and check clinical areas to see if any remain in use. Consignees are to ensure that all personnel, including clinical staff, are completely knowledgeable of this recall notification and addendum, which instructs users to slide the endotracheal tube along the entire length of the tracheal intubation scope insertion section to confirm compatibility before using on a patient. Consignees are asked to complete and return the provided response form to Sedgwick via email at Olympus7498@sedgwick.com or by fax at 866-552-4917. Any questions about the acknowledgement form may be directed by phone to 866-737-4762. If product has been further distributed, please forward the notification to those secondary consignees and maintain records of your documentation process. Consignees are to report any complaints to Olympus Technical Assistance Center at 1-800-848-9024 (Option 1) and FDA Medwatch. Any further questions are to be directed to Tara Safi by email at Tara.Safi@Olympus.com.
Root cause
Labeling Change Control
Status
Open, Classified
Product code
EOQ
Quantity affected
3949 units
Distribution
US Nationwide distribution.
Date initiated
2023-07-27
Date posted
2023-08-31
Location
Center Valley, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
EVIS EXERA III Bronchovideoscope Olympus BF-XP190, EVIS EXERA III Bronchovideoscope Olympus BF-P190 (K201758)Olympus Medical Systems Corp.2021-01-21