OEC Elite: The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy

FDA device recall Z-1819-2025 · posted 2025-06-09 · Open, Classified

Recall details

Recalling firm
GE Medical Systems, LLC
Reason for recall
OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
Action taken
All impacted customers with receive a customer letter notifying them of the potential issue. GE HealthCare will send a field service engineer to perform measurements of the Xray field on the identified systems (listed in Appendix A of the customer letter).
Root cause
Process control
Status
Open, Classified
Product code
OXO
Quantity affected
Total for both products: 3472 (2,919 US; 553 OUS)
Distribution
US and OUS
Date initiated
2025-03-21
Date posted
2025-06-09
Location
Waukesha, WI