OBSERVA R02 (version R02.00.17) computers using PSC 6000 or HHP Barcode Scanners which are connected to BacT/ALERT 3D

FDA device recall Z-1161-06 · posted 2006-06-22 · Terminated

Recall details

Recalling firm
bioMerieux Inc
Reason for recall
Software problem with scanner inter-character delay which may result in false positive or false negative results after incorrect scanning of bottle ID's
Action taken
The recalling firm telephoned customers who ordered the MB bottles and use the OBSERVA software ver. 2.0 to inform them of the issues. Recall letters were issued to all customers via overnight mail on 5/22/06 informing them a field representative would be visiting to make a short-term correction to the system. The letter also informed the customer a permanent correction to the software was scheduled for release later in the year.
Root cause
Other
Status
Terminated
Product code
MDB
Quantity affected
51 units
Distribution
WorldwideDistribution--Nationwide to medical centers. Government distribution was made to MN and CA, and military distribution was made to OK. Foreign distribution was made to Thailand, The Netherland
Date initiated
2006-03-28
Date posted
2006-06-22
Date terminated
2008-03-07
Location
Hazelwood, MO