O&M Halyard Procedure Mask with SO SOFT* Earloops, Yellow-Disposable surgical mask Product Code: 47117

FDA device recall Z-0728-2025 · posted 2024-12-20 · Open, Classified

Recall details

Recalling firm
O&M HALYARD, INC.
Reason for recall
Not meeting the bioburden testing requirements of EN 14683 required for masks sold in Europe. Face masks sold in the United States are not required to meet this standard. However, if the standard is printed on the packaging, compliance is required.
Action taken
O&M Halyard issued Urgent: Product Field Action via email on 11/8/24 to Distributors and End-Users. Letter states reason for recall, health risk and action to take: Customers (end-users): 1. REVIEW your inventory for the affected product code and specific lot number. 2. DESTROY or DISCARD the product per the facility s standard disposal procedure for face masks. 3. COMMUNICATE this Field Action to staff to ensure awareness of it. 4. RETURN the completed Field Action Response Form to OMRA_RECALLS@owens-minor.com as soon as possible even if you have no current product inventory. 5. Any adverse events associated with the use of the affected product should be reported to complaints@owens-minor.com. 6. If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. Distributors: 1. REVIEW your inventory for the affected product code and specific lot number. 2. DESTROY or DISCARD the product per the facility s standard disposal procedure for face masks. 3. COMMUNICATE this voluntary Field Action to staff to ensure awareness of it. 4. If any of the affected lot identified above was further distributed, please ensure all end-users are appropriately notified of this Field Action and maintain records of effectiveness. 5. Any adverse events associated with the use of the affected product should be reported to complaints@owens-minor.com. 6. RETURN the completed Field Action Response Form to OMRA_RECALLS@owens-minor.com as soon as possible even if you have no current product inventory. If you have any questions, please email OMRA_RECALLS@owens-minor.com using the Event # FA-2024-034 in the header of the response.
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
FXX
Quantity affected
211 cases
Distribution
Nationwide
Date initiated
2024-11-08
Date posted
2024-12-20
Location
Mechanicsville, VA