Atlas FDA

O-Arm Imaging System The O-Arm imaging system is a mobile X-Ray system designed for medical applications in surgical the

FDA device recall Z-1998-2010 · posted 2010-07-13 · Terminated

Recall details

Recalling firm
Medtronic Navigation, Inc.
Reason for recall
The possibility that the real time image display can freeze up during use of a 2D continuous fluoroscopy.
Action taken
The firm, Medtronic, Inc., sent an "Urgent Field Safety Notice" letter dated May 10, 2010, to all custmers. The letter describes the product, problem and actions to be taken by customer. The firm recommended that the users of the O-Arm Imaging System reveiw the information provided to ensure patient safety remains uncompromised during use of the product. The customers were instructed to discontinue use and contact your local Medtronic Navigation representative or call our Technical Services Group at 800-595-9709 or (702) 890-3200, if the image is not updating as expected. A Medtronic Navigation representative will be contacting you within the coming months to schedule a time to evaluate and service the potentially affected components within your system. A service visit will also be scheduled starting November 2010 to update your system with a software based solution that will prevent this issue from occuring. If you have any questions about this notification, please contact your local Medtronic Navigation representative or call our Technical Services Group at 800-595-9709 or (702) 890-3200.
Root cause
Device Design
Status
Terminated
Product code
LLZ
Quantity affected
221
Distribution
Worldwide distribution: USA and countries of Germany, Singapore, South Africa, Poland, Libya, Belgium, Switzerland, Turkey, Italy, Denmark, France, Spain, Finland, Czech Republic, Ireland, Slovakia, t
Date initiated
2010-05-10
Date posted
2010-07-13
Date terminated
2011-10-07
Location
Littleton, MA