NutraGlide Nasal Feeding Tube with Stylet and ENFit, 5 F - 8 F x 55, REF E NRT-06055-I The NutraGlide Nasal Feeding Tube
FDA device recall Z-0129-2026 · posted 2025-10-09 · Completed
Recall details
- Recalling firm
- Applied Medical Technology Inc
- Reason for recall
- The nasal feeding tubes, NutraGlide, subject to this recall have distal tips that may detach at lower than expected forces.
- Action taken
- AMT issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 8/12/2025 via USPS certified mail. The notice explained the issue, risk to health and requested the following: ACTIONS TO BE TAKEN: CUSTOMER OR USER: Applied Medical Technology, Inc. requests that customers IMMEDIATELY TAKE THE FOLLOWING ACTIONS: " Examine your inventory and isolate/quarantine pending return of the product subject to this recall. " Do not use any device subject to this recall. " Complete and return to Applied Medical Technology, Inc. the enclosed RECALL NOTIFICATION ACKNOWLEDGMENT via fax or send by mail. " Please contact Applied Medical Technology, Inc. either by phone at 440-717-4000, or via email at cs@appliedmedical.net, to make arrangements for the return of any devices related to this recall. Include the product, all of its components, and any available packaging and labeling associated with the device.
- Root cause
- Process control
- Status
- Completed
- Product code
- KNT
- Quantity affected
- 840 units
- Distribution
- US distribution to MA and RI
- Date initiated
- 2025-08-12
- Date posted
- 2025-10-09
- Location
- Brecksville, OH