NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instrument).
FDA device recall Z-0735-04 · posted 2004-07-20 · Terminated
Recall details
- Recalling firm
- Biomerieux, Inc.
- Reason for recall
- Instrument malfunction. Electrical short circuit may occur resulting in damage or a fire hazard.
- Action taken
- This correction is being handled as an instrument repair (rework) by bioMerieux Field Service; therefore no communication has been sent to the customer sites. The repair involves heat sink replacement or use of insulating tape on the rear side of the keypad.
- Root cause
- Other
- Status
- Terminated
- Product code
- JJQ
- Quantity affected
- 3 units
- Distribution
- CT, MD, MT
- Date initiated
- 2004-02-26
- Date posted
- 2004-07-20
- Date terminated
- 2004-03-24
- Location
- Durham, NC