NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instrument).

FDA device recall Z-0735-04 · posted 2004-07-20 · Terminated

Recall details

Recalling firm
Biomerieux, Inc.
Reason for recall
Instrument malfunction. Electrical short circuit may occur resulting in damage or a fire hazard.
Action taken
This correction is being handled as an instrument repair (rework) by bioMerieux Field Service; therefore no communication has been sent to the customer sites. The repair involves heat sink replacement or use of insulating tape on the rear side of the keypad.
Root cause
Other
Status
Terminated
Product code
JJQ
Quantity affected
3 units
Distribution
CT, MD, MT
Date initiated
2004-02-26
Date posted
2004-07-20
Date terminated
2004-03-24
Location
Durham, NC