Nova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Cr

FDA device recall Z-0650-2014 · posted 2014-01-09 · Terminated

Recall details

Recalling firm
Nova Biomedical Corporation
Reason for recall
Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage
Action taken
On November 18, 2013, an URGENT Field Correction Letter was sent to all StatStrip Glucose Hospital meter, StatStrip Lactate Hospital Meter and StatSensor Creatinine Hospital Meter customers via email to the Point of Care Coordinator to alert users that lithium batteries that are utilized beyond their intended useful life may swell and/or crack posing a potential safety hazard. In some cases, the swelling may potentially cause the battery door on the meter to dislodge and render the meter inoperable.
Root cause
Component design/selection
Status
Terminated
Product code
CGA
Quantity affected
est. 200,000
Distribution
Nationwide Distribution.
Date initiated
2013-10-10
Date posted
2014-01-09
Date terminated
2014-03-24
Location
Waltham, MA