Atlas FDA

Nova Max Glucose Test Strips (50 Count International Catalog Number: 45512 Product usage: The Nova Max Glucose Test Stri

FDA device recall Z-1902-2013 · posted 2013-08-09 · Terminated

Recall details

Recalling firm
Nova Biomedical Corporation
Reason for recall
Reports of false abnormally high glucose readings from test strips
Action taken
Nova Biomedical U.S.A issued an Urgent Medical Device Voluntary Recall Letter dated July 19, 2013 via registered/certified letters to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately discontinue use of the affected product; contact Nova Diabetes Care Customer Service for replacement strips; do not stop testing blood glucose while waiting for replacement strips; if affected strips must be used follow precautions provided in notification letter. For questions call Nova Diabetes Care Customer Service at 1-800-681-7390. .
Root cause
Material/Component Contamination
Status
Terminated
Product code
NBW
Quantity affected
366,140 test strips
Distribution
Worldwide Distribution - USA Nationwide and the countries of Chile, England , Peru, Nicaragua
Date initiated
2013-07-15
Date posted
2013-08-09
Date terminated
2017-08-01
Location
Waltham, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
NOVA MAX PLUS BLOOD GLUCOSE AND B-KETONE MONITOR SYSTEM, NOVA MAX PLUS B-KETONE CONTROL SOLUTIONS, NOVA MAX BLOOD GLUCOS (K091547)Nova Biomedical Corp.2010-01-15