Nova 8 Analyzer Calibrator Pack (Catalog/REF # 15196), Nova Biomedical, Waltham, MA
FDA device recall Z-0295-2008 · posted 2008-01-17 · Terminated
Recall details
- Recalling firm
- Nova Biomedical Corporation
- Reason for recall
- Elevated Normalized Ionized Calcium (nCa) and Normalized Ionized Magnesium (nMg) calculated values on patient samples
- Action taken
- Nova issued a Customer Advisory Notice notification by phone on 9/17/07 to contact customers using a script . If the facility reports Normalized Ionized Calcium or Normalized Ionized Magnesium, Nova Biomedical will replace existing stock of Nova 8 calibrator packs. If the facility does not report Normalized Ionized Calcium or Normalized Ionized Magnesium, the account can continue to use the existing lot of calibrator packs.This contact will be documented within the Nova Biomedical complaint database. Intemational Customer Advisory Notice being sent to Nova Biomedical Subsidiariies and Distributors by Federal Express Envelope.The Federal Express envelopes tracking number will be used for delivery confirmation.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- JBS
- Quantity affected
- 2,999 units
- Distribution
- Worldwide, USA, Canada, Argentina, Bangladesh, Bolivia, China, Czech Republic, Denmark, Ecuador, Egypt, Finland, France,Germany,Greece, Hungary, India, Indonesia, Ireland,Italy, Japan, Korea, Mexico,
- Date initiated
- 2007-09-17
- Date posted
- 2008-01-17
- Date terminated
- 2012-03-16
- Location
- Waltham, MA