Nova 8 Analyzer Calibrator Pack (Catalog/REF # 15196), Nova Biomedical, Waltham, MA

FDA device recall Z-0295-2008 · posted 2008-01-17 · Terminated

Recall details

Recalling firm
Nova Biomedical Corporation
Reason for recall
Elevated Normalized Ionized Calcium (nCa) and Normalized Ionized Magnesium (nMg) calculated values on patient samples
Action taken
Nova issued a Customer Advisory Notice notification by phone on 9/17/07 to contact customers using a script . If the facility reports Normalized Ionized Calcium or Normalized Ionized Magnesium, Nova Biomedical will replace existing stock of Nova 8 calibrator packs. If the facility does not report Normalized Ionized Calcium or Normalized Ionized Magnesium, the account can continue to use the existing lot of calibrator packs.This contact will be documented within the Nova Biomedical complaint database. Intemational Customer Advisory Notice being sent to Nova Biomedical Subsidiariies and Distributors by Federal Express Envelope.The Federal Express envelopes tracking number will be used for delivery confirmation.
Root cause
Device Design
Status
Terminated
Product code
JBS
Quantity affected
2,999 units
Distribution
Worldwide, USA, Canada, Argentina, Bangladesh, Bolivia, China, Czech Republic, Denmark, Ecuador, Egypt, Finland, France,Germany,Greece, Hungary, India, Indonesia, Ireland,Italy, Japan, Korea, Mexico,
Date initiated
2007-09-17
Date posted
2008-01-17
Date terminated
2012-03-16
Location
Waltham, MA