Normed General Instruments 503004541: Werber Countersink Cannulated for micro CBS Screws, AO 503004341: Werber Countersi
FDA device recall Z-2111-2017 · posted 2017-05-15 · Terminated
Recall details
- Recalling firm
- Zimmer Gmbh
- Reason for recall
- the devices used a different raw material to produce two countersink products which has higher hardness and higher carbon content resulting in device damage.
- Action taken
- Zimmer Biomet sent an URGENT MEDICAL DEVICE RECALL REMOVAL letter dated May 11 2017. Lot Specific notifications were sent to the affected consignee via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions were instructed to call Technical Services at 1-574-371-3071.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- HTT
- Quantity affected
- 138 (OUS 297)
- Distribution
- Single US consignee CO.
- Date initiated
- 2017-05-11
- Date posted
- 2017-05-15
- Date terminated
- 2017-11-03
- Location
- Winterthur, Switzerland