Atlas FDA

Normed General Instruments 503004541: Werber Countersink Cannulated for micro CBS Screws, AO 503004341: Werber Countersi

FDA device recall Z-2111-2017 · posted 2017-05-15 · Terminated

Recall details

Recalling firm
Zimmer Gmbh
Reason for recall
the devices used a different raw material to produce two countersink products which has higher hardness and higher carbon content resulting in device damage.
Action taken
Zimmer Biomet sent an URGENT MEDICAL DEVICE RECALL REMOVAL letter dated May 11 2017. Lot Specific notifications were sent to the affected consignee via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with questions were instructed to call Technical Services at 1-574-371-3071.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
HTT
Quantity affected
138 (OUS 297)
Distribution
Single US consignee CO.
Date initiated
2017-05-11
Date posted
2017-05-15
Date terminated
2017-11-03
Location
Winterthur, Switzerland