NobelParallel Conical Connection NP, 3.75x10mm (Endosseous Dental Implant); REF 37965; UDI (01)07332747091441(10)1212910

FDA device recall Z-1851-2020 · posted 2020-05-04 · Terminated

Recall details

Recalling firm
Nobel Biocare Usa Llc
Reason for recall
The top label of article no. 37965, lot 12129104 (immediate packaging of endosseous dental implants ) may incorrectly indicate an implant length of 7mm, which is shorter than the actual length of the implant (10mm).
Action taken
The firm's "Urgent Field Safety Notice", was issued on 08/28/2019 and the notices were sent to customers via email . The following actions are to be taken: 1) Identify affected devices in stock by confirming the information on the side label: 3.75x10mm, article no. 37965, lot 12129104. 2) Confirm that both the side and top label indicate 3.75x10mm. 3) Complete a Customer Acknowledgment Form and return it via email to: cs.russia@nobelbiocare.com within 5 days of receipt of the notice. 4) Return incorrectly labeled products (top label 3.75x7mm) to: Attn: Returns PFA1925 109004 Moscow Stanislavskogo str., h.21, b.3 (Correctly labeled products (top label 3.75x 10mm) do not need to be returned). 5) Returned devices will be replaced free of charge.
Root cause
Labeling mix-ups
Status
Terminated
Product code
DZE
Quantity affected
228 pieces
Distribution
International distribution - the country Russia only.
Date initiated
2019-08-28
Date posted
2020-05-04
Date terminated
2021-04-28
Location
Yorba Linda, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
NobelParallel Conical Connection (K173418)Nobel Biocare AB2019-02-21