NicoletOne photic adapter cable 085-463700

FDA device recall Z-2389-2008 · posted 2008-09-21 · Terminated

Recall details

Recalling firm
Cardinal Health NeuroCare Division
Reason for recall
Cardinal Health NeuroCare is implementing a voluntary field correction to replace a quantity of photic adapter cables, part number 085-463700, used to connect the NicLED Photic Strobe option, part number 698-625500, with NicoletOne EEG system. The photic adapter cable has a wiring error which delivers a reduced voltage and results in a decreased intensity of the NicLED photic. This reduced intensi
Action taken
A "Cardinal Health Urgent: Medical Device Correction" letter was sent to consignees. The letter described the product and the problem. The consignees were asked to confirm completion of the cable exchange by returning an information card. If you have questions, contact Cardinal Health (formerly VIASYS) NeuroCare Technical Support Group at 1-800-356-0007 (US), 608-273-5000 (OUS) or email shane.sawall@cardinalhealth.com.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
GWQ
Quantity affected
92
Distribution
CA FL GA ID IN LA MD ME MI MO NC NY PA TN TX WA WI China Denmark Indonesia Japan Korea Mexico Netherlands Russia Saudi Arabia Slovenia
Date initiated
2008-06-05
Date posted
2008-09-21
Date terminated
2010-10-13
Location
Madison, WI