Atlas FDA

NICO Myriad Console Model number NN-7000 The Myriad System is a surgical system specifically designed for morcellation o

FDA device recall Z-2767-2015 · posted 2015-09-15 · Terminated

Recall details

Recalling firm
Nico Corp.
Reason for recall
Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.
Action taken
The firm, Nico Corporation, sent an "URGENT: NICO Console Recall" notification dated 7/31/15 via fax to its customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The customers were instructed to return the console and complete and return the Recall Response Form via fax to: 317-683-0305 or email to: Jay.Dittman@niconeuro.com. The firm will provide customers with a Myriad Console loaner kit. If you have any questions, contact Director of Quality and Regulatory, at 317-660-7118 ext 104 or email to: jay.dittman@nicneuro.com.
Root cause
Process control
Status
Terminated
Product code
GEY
Quantity affected
6 units
Distribution
US Distribution to states of: MD, IN, CA, and NC.
Date initiated
2015-07-29
Date posted
2015-09-15
Date terminated
2015-09-22
Location
Indianapolis, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SURGICAL CUTTER (K955168)Promex, Inc.1996-01-22