Atlas FDA

NexGen Complete Knee Solution Stemmed tibial broach impactor, Catalog # 00-5977-011-00, Zimmer Inc., Warsaw, IN.

FDA device recall Z-0609-2011 · posted 2010-12-13 · Terminated

Recall details

Recalling firm
Zimmer Inc.
Reason for recall
The impactor may disassemble or detach during use. The detached piece may not be observed intraoperatively and may be left in the surgical site. Complications may include pain, infection, and prolongation in surgery. These complications may also require revision surgery.
Action taken
Zimmer issued "Urgent Medical Device Safety Alert" letters dated April 16, 2010 to Risk Managers, user facilities, Health Care Professionals and Zimmer Sales force. The users were informed to inspect the device for loosening. The alert also informed users to inspect the device after each use to verify that the device was intact. Customers were instructed to complete and return a User Facility Documentation Form. Zimmer will repair and replace all affected devices. Customers can contact ZImmer, Inc. at 800 613-6131 concerning this recall.
Root cause
Device Design
Status
Terminated
Product code
LXH
Quantity affected
10,666 all products.
Distribution
Worldwide Distribution: Throughout US including Puerto Rico, and the countries of Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Denmark, Ecuador, El Salvador, France,
Date initiated
2010-04-16
Date posted
2010-12-13
Date terminated
2012-09-12
Location
Warsaw, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
NEXGEN COMPLETE KNEE SOLUTION LPS-FLEX AND KNEE GENDER SOLUTIONS FEMALE (GSF) POROUS FEMORAL COMPONENTS (K072619)Zimmer, Inc.2007-11-21
NEXGEN COMPLETE KNEE SOLUTION LPS-FLEX PROLONG HIGHLY CROSSLINKED POLYETHYLENE ARTICULAR SURFACES (K042271)Zimmer, Inc.2004-10-13
NEXGEN COMPLETE KNEE SOLUTION LEGACY POSTERIOR STABILIZED (LPS); LPS-FLEX FIXED BEARING FEMORAL AND ARTICULAR SURFACE CO (K991581)Zimmer, Inc.1999-07-30
NEXGEN COMPL. KNEE SOL. LEGACY POSTERIOR STABILIZED/CONSTRAINED CONDYLAR KNEE FEMORAL COMPONENTS/ARTICULAR SURFACES (K960279)Zimmer, Inc.1996-04-26
NEXT GENERATION KNEE SYSTEM (K933785)Zimmer, Inc.1995-01-30