Atlas FDA

NeuViz Dual series Computed Tomography (CT) Scanner System, X-Ray System. Part Number: 989605651321. Intended to produce

FDA device recall Z-2237-2010 · posted 2010-08-19 · Terminated

Recall details

Recalling firm
Neuisys, LLC
Reason for recall
The reference lines for image generated for a Surview scan may appear in the incorrect position in the "FILM" display mode.
Action taken
On 07/21/2008, NEUISYS Imaging Systems Solutions distributor in Greensboro, NC began sending the URGENT DEVICE CORRECTION letter on behalf of Philips and NeuSoft Medical Systems (PNMS), Co., Ltd in China. The letter informs the consignees that there is potential safety problem in the NeuViz Dual series CT scanners due to the malfunction of "Incorrect Reference Line in the Film". When a "Head first, Couch out" or a "Feet first, Couch in" Surview scan is started then paused midstream, the image generated is what the doctor expects, but the reference lines appear in the incorrect position in "FILM" display mode. The acquired images accurately represent the exact positions planned on the original view. The incorrect/unmatched reference line in film may result in incorrect diagnosis. The consignees are advised not to stop or pause during Surview acquisition while PNMS in China will provide the updated software patch to all affected systems free of charge. The Field Change Order #FCOP-08-CT-NMS-008 SP4 will be implemented by NEUISYS Imaging Systems Solutions in Greensboro, NC., who is a distributor for PNMS. Consignees are advised to e-mail Service Support Department of PNMS at helpdesk@pnms.neusoft.com regarding the recall.
Root cause
Software design
Status
Terminated
Product code
JAK
Quantity affected
19 units in the US
Distribution
Nationwide Distribution
Date initiated
2008-07-21
Date posted
2010-08-19
Date terminated
2010-08-23
Location
Greensboro, NC

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
MODIFICATION TO NEUVIZ DUAL MULTI-SLICE CT SCANNER SYSTEM (K071308)Neusoft Medical Systems Co., Ltd.2007-08-16
NEUVIZ DUAL MULTI-SLICE CT SCANNER SYSTEM (K062451)Neusoft Medical Systems Co., Ltd.2006-09-28