NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body computed tomography X-ray system featuring a conti

FDA device recall Z-1573-2016 · posted 2016-04-27 · Terminated

Recall details

Recalling firm
Philips and Neusoft Medical Systems Co., Ltd.
Reason for recall
The following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426: 1) During the filming operation on MX 16-slice console software, the clipboard used for copying and pasting images is not cleared between patients. If the operator fails to copy the current patient's images before pasting, a previous patient's image may be present in the clipboard and be
Action taken
Affected consignees will be notified of the recall via letter on March 24, 2016. Mandatory Field Change Order and FCO kits will be released to all affected systems in June, 2016.
Root cause
Under Investigation by firm
Status
Terminated
Product code
JAK
Quantity affected
37 units
Distribution
NC, OH, NE, SC, TX, LA, PR, MO, FL, CT
Date initiated
2016-03-24
Date posted
2016-04-27
Date terminated
2016-11-23
Location
Sujiatun Shenyang, China