Atlas FDA

NeuViz 16 Multi-Slice CT Scanner System is a computed Tomography X-ray system, and has a Part number 989605858501. The d

FDA device recall Z-2083-2012 · posted 2012-07-26 · Terminated

Recall details

Recalling firm
Philips And Neusoft Medical Systems Co., Ltd.
Reason for recall
One of the four screws that secure the Oil-accumulator to the X-Ray Tube pulled out in the NeuViz 16 CT scanner, which caused the Oil-accumulator to detach.
Action taken
Philips Healthcare North America Company sent an "Urgent Device Correction NeuViz 16 X-Ray Tube Oil-accumulator Detached" letter and attachments of Annex A Method to Identify Affected Products and Annex B Instruction for Mechanic, dated June 14, 2012, to all affected customers. The letter described the product, problem, and actions to be taken by the customers. Customers are advised to follow the instructions on Annex A and Annex B. Philips and Neusoft Medical Systems planned to release the mandatory Field Change Order (FCO) and FCO kits to all customers by the end of June 2012. If you have any questions about this corrective action program, please contact the Service Support Department, Neusoft Medical Systems, Co., LTD by e-mail to NMS-SERVICE@NEUSOFT.COM
Root cause
Device Design
Status
Terminated
Product code
JAK
Quantity affected
15 units distributed in the US
Distribution
Nationwide Distribution -- North Carolina, Ohio, and Texas.
Date initiated
2012-06-08
Date posted
2012-07-26
Date terminated
2012-09-17
Location
Shenyang, Liaoning, China

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
NEUVIZ 16 MULTI-SLICE CT SCANNER SYSTEM (K092742)Neusoft Medical Systems Co., Ltd.2009-09-22
NEUVIZ 16 MULTI-SLICE CT SCANNER SYSTEM (K090173)Neusoft Medical Systems Co., Ltd.2009-02-05