Atlas FDA

Needle Counters; 11.5 x 6.5 x 3.3 cm; 20 Count/40 Capacity; Sterile; Removable Numbered Foam Block; Adhesive Area (Cat.

FDA device recall Z-0978-2013 · posted 2013-03-28 · Terminated

Recall details

Recalling firm
Cardinal Health
Reason for recall
Small holes (approximately 1 mm or less) found in device unit level packaging potentially compromising product sterility.
Action taken
Cardinal Health sent an Urgent Product Recall letter dated December 18, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immedaitely stop using affected product and discard any product that may be on hand, return the enclosed acknowledgement form, notify any customers to whom you may have distributed product affected by this recall, and contact the appropriate Customer Service group to arrange for credit of any affected product discarded. For questions contact Cardinal Health Professional Services at 800-292-9332.
Root cause
Package design/selection
Status
Terminated
Product code
FSM
Quantity affected
Cat. No. 3FA40SSA: 23,952; Cat. No. 3FM40SSA: 140,736
Distribution
Worldwide Distribution - USA (Nationwide) and the countried of Australia, Canada and Japan.
Date initiated
2012-12-18
Date posted
2013-03-28
Date terminated
2016-08-01
Location
Park, IL