NC Tenku RX PTCA Balloon Catheter, Part No. 1012448-08T, 1012448-12T, 1012448-15T, 1012449-08T, 1012449-12T, 1012449-15T
FDA device recall Z-2039-2017 · posted 2017-05-12 · Terminated
Recall details
- Recalling firm
- Abbott Vascular
- Reason for recall
- Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon.
- Action taken
- An Urgent Field Safety Notice dated 3/22/17 was sent to customers to inform them that Abbott Vascular is recalling the NC Trek RX Coronary Dilation Catheter, NC Traveler RX Coronary Dilatation Catheter, and NC Tenku RX PCTA Balloon Catheter. Products from the identified lots may exhibit difficulty in removing the protective balloon sheath which can result in issues with inflating or deflating the balloon. The letter informs the customers of how does the issue occur and what action is Abbott Vascular asking the customers to take. Customers with any questions are instructed to contact local Abbott Vascular Representative or Customer Service Department at (800) 227-9902. On 5/15/17, Abbott Vascular posted a press release on their website to inform customers that Abbott has initiated a voluntary recall of specific lots of three catheters: NC Trek RX Coronary Dilatation Catheter, NC Traveler Coronary Dilatation Catheter, and NC Tenku RX PTCA Balloon Catheter. Products from the identified lots may exhibit difficulty in removing the protective balloon sheath, which could cause problems with inflating or deflating the balloon. Potential risks associated with balloon inflation and deflation difficulties include air embolism, additional intervention, thrombosis, and myocardial infarction.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- LOX
- Quantity affected
- 449,661 units total (132,040 units in US)
- Distribution
- US and Worldwide
- Date initiated
- 2017-03-22
- Date posted
- 2017-05-12
- Date terminated
- 2018-02-07
- Location
- Temecula, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| VOYAGER NC CORONARY DILATATION CATHETER AND NC TREK RX CORONARY DILATATION CATHETER (K103153) | Abbott Vascular-Cardiac Therapies | 2010-12-30 |