Navilyst Medical, Inc., Vaxcel(TM) Port Plastic Standard Port with 8F Silicone Catheter (1.5 mm/2.6 mm), REF Catalog No.
FDA device recall Z-1886-2010 · posted 2010-07-01 · Terminated
Recall details
- Recalling firm
- Navilyst Medical, Inc
- Reason for recall
- FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.
- Action taken
- Navilyst Medical, Inc. is recalling Vaxcel Ports and Vaxcel Ports with PASV Valve with valid expiration dates worldwide. Navilyst sent "Urgent Medical Device Recall - Immediate Action Required" letter/recall notification packages on May 13, 2010, via Federal Express to 471 domestic consignees; on May 14, 2010, via e-mail to 14 international consignees and to Boston Scientific Corporation (prior owner of the affected products) via facsimile and Federal Express. On June 3, 2010, Navilyst Medical sent Recall Notification Packages on behalf of Boston Scientific Corp. via Federal Express to 182 domestic consignees. The letter instructs the customer to immediately discontinue use of and segregate recalled product, complete and return the reply verification tracking form to fax# 518-742-4450 and package and return the recalled product to the address provided. If you have any question, please call Navilyst Medical, Inc., Customer Service at 1-800-833-9973 or Michael Duerr at 1-518-742-4571.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- LJT
- Distribution
- Worldwide distribution:USA and countries of Belgium, Brazil, Canada, Colombia, Croatia, Germany, Greece, Hong Kong, Italy, Saudi Arabia, South Korea, Spain, Tunisia, and United Kingdom.
- Date initiated
- 2010-05-13
- Date posted
- 2010-07-01
- Date terminated
- 2012-08-03
- Location
- Glens Falls, NY
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| TRIMPORT IMPLANTABLE ACCESS SYSTEM (MODIFICATION) (K954258) | Gerard Medical Enterprises, Inc. | 1996-05-24 |
| TRIMPORT (K945250) | Gerard Medical Enterprises, Inc. | 1995-03-07 |
| TRIMPORT (K942623) | Gerard Medical Enterprises, Inc. | 1994-08-23 |
| GERARD MEDICAL, INC. TRIM-PORT (K903364) | Gerard Medical Enterprises, Inc. | 1990-10-31 |