Atlas FDA

NaturaLyte Liquid Bicarbonate Concentrate (Dialysate Concentrate for Hemodialysis (liquid), 6.4 Liter Bottle Catalog Num

FDA device recall Z-1827-2015 · posted 2015-06-22 · Open, Classified

Recall details

Recalling firm
Fresenius Medical Care Holdings, Inc.
Reason for recall
Bacterial contamination.
Action taken
Fresenius Medical Care sent an Urgent Medical Device Recall letter dated May 15, 2015, to all affected customers. Users were requested to Immediately examine stock to determine whether they have any Naturalyte Liquid Bicarbonate Concentrate of the lots. " If any product of these lots is found, discontinue use immediately. " Place all units in a secure, segregated area. " If affected product was on the machine prior to patient treatment, perform a [Heat Disinfect] program. " Your dialysis schedule should not be interrupted. If interruption of your dialysis schedule is expected, please discuss your options with your health care provider. " Contact FMCNA Customer Service Team at 1-800-323-5188 for instructions on how to return the recalled product. " Promptly fill out and return the attached reply form Additional medical concerns or questions, please contact Medical Information and Communication: 855-616-2309 or Website: www.fresenius-medinfo.com. For questions regarding this recall call 800-662-1237.
Root cause
Process control
Status
Open, Classified
Product code
KPO
Quantity affected
1,856,619 Bottles
Distribution
Nationwide Distribution
Date initiated
2015-05-20
Date posted
2015-06-22
Location
Waltham, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
FRESENIUS NATURALYTE SODIUM BICARBONATE LIQUID CONCENTRATE, MODEL 08-4000-LB (K071387)Fresenius Medical Care North America2007-06-15