Atlas FDA

nanOss 3D Advanced Bone Graft Substitute nanOss 3D Plus Advanced Bone Graft Substitute 90-300-251008: nanOss 3D Advanced

FDA device recall Z-1897-2026 · posted 2026-04-21 · Open, Classified

Recall details

Recalling firm
XTANT Medical Holdings, Inc
Reason for recall
Products were incorrectly labeled with a 5-year shelf life. The approved and validated shelf life is four years.
Action taken
On March 9, 2026, the firm began notifying customers via letters titled "URGENT: MEDICAL DEVICE RECALL." Customers were instructed to segregate and quarantine all affected products to prevent use, and discontinue use of any affected product. Affected product must be returned to Xtant Medical.
Root cause
Incorrect or no expiration date
Status
Open, Classified
Product code
MQV
Quantity affected
225
Distribution
US Nationwide distribution in the states and territories of CA, CT, DC, KS, MD, NC, NY, TX, and WI.
Date initiated
2026-03-09
Date posted
2026-04-21
Location
Belgrade, MT

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
NB3D BONE VOID FILLER (K132050)Pioneer Surgical Technology, Inc.2014-02-07