Atlas FDA

NanoDots, Model Numbers: a) BC30023, description: QC Set For Microstar Dots; b) BC30083, description: nanoDot QC Set - 8

FDA device recall Z-2573-2023 · posted 2023-09-12 · Completed

Recall details

Recalling firm
Landauer
Reason for recall
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Action taken
Landauer disseminated an URGENT - Medical Device Recall notice to its consignees on 07/12/2023 by US mail and email. The notice explained the problem with the devices, potential risk, and requested the following: Discontinue use of the affected products, destroy used products, and return unused products.
Root cause
Under Investigation by firm
Status
Completed
Product code
IYE
Quantity affected
18333 units
Distribution
Worldwide
Date initiated
2023-07-12
Date posted
2023-09-12
Location
Glenwood, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Medical Radiation Dosimetry System microSTARii (K192196)Landauer2019-12-27