namic convenience kits labeled as: MTS,COLOR,PML,KIT,-,FRENCH HOSPITAL PG; medical convenience kit, REF 600605710
FDA device recall Z-1430-2025 · posted 2025-03-21 · Open, Classified
Recall details
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Reason for recall
- The female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.
- Action taken
- Medline Industries issued two notices, delivered using email and US Mail: 1) a MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines and 2) a MEDICAL DEVICE RECALL notice for the kits containing the affected product; both notices explained the issue and the hazard. The MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines requested that the consignee destroy any affected inventory and reply to receive credit. The MEDICAL DEVICE RECALL notice for the kits containing the affected product requested the consignee respond to the notice and upon response, Medline will provide labeling that states the following: "ATTENTION This kit contains a recalled item Medline Pressure Monitoring Lines VPN: 24614303 DO NOT USE THIS COMPONENT The rest of the kit may be used. Please remove this item only and replace with product from supply. *Do not cover or remove this sticker*" Consignees were directed to notify their customers of these actions.
- Root cause
- Process change control
- Status
- Open, Classified
- Product code
- OEQ
- Quantity affected
- 100 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia
- Date initiated
- 2025-02-12
- Date posted
- 2025-03-21
- Location
- Northfield, IL