Myoscience brand iovera Smart Tip, Myoscience Cryo-Touch IV, a Cryogenic Surgical device, REF Catalog Number: STA0311-10

FDA device recall Z-1781-2013 · posted 2013-07-20 · Terminated

Recall details

Recalling firm
Myoscience Inc
Reason for recall
The outer box label of the product was labeled with a down revision label which did not include the US labeling requirements, specifically the symbol descriptors and also the "Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician." This product was previously labeled for EU distribution only.
Action taken
Myoscience representatives visited both affected sites on June 14, 2013 and relabeled the product in person. Neither site had used the product and was unaffected by the relabeling. No formal communication was issued.
Root cause
Labeling mix-ups
Status
Terminated
Product code
GXH
Quantity affected
3 units
Distribution
US distribution including CA and ID.
Date initiated
2013-06-21
Date posted
2013-07-20
Date terminated
2014-07-15
Location
Redwood City, CA