MYLA server HP Proliant: DL380-G8 and ML350-G6 version 3.xx computer application software. MYLA V3 is a computer applica

FDA device recall Z-1621-2015 · posted 2015-05-15 · Terminated

Recall details

Recalling firm
bioMerieux, Inc.
Reason for recall
The MYLA server could slow down due to the volume (weight) of the data to manage and it could also stop processing, impacting the server start-up phase. The problem occurs in high volume settings. The MYLA server slows down and will stop processing completely if any task requires more than 5 minutes to complete. The server then fails to reboot and requires service, provided remotely or on-site
Action taken
Consignees were notified electronically on April 14, 2015 and letter via FedEx on 04/20/2015.
Root cause
Software design
Status
Terminated
Product code
N/A
Quantity affected
730 units
Distribution
Worldwide Distribution-Algeria, Argentina, Australia, Austria, Brazil Canada, Chile China, Colombia, Czech Republic, France Germany , Greece, Hungary, India, Italy Japan, Korea, Mexico, Netherlands, P
Date initiated
2015-04-14
Date posted
2015-05-15
Date terminated
2017-12-13
Location
Durham, NC