Atlas FDA

Multi Joystick R-net. Electrical wheelchair component.

FDA device recall Z-2455-2025 · posted 2025-09-02 · Open, Classified

Recall details

Recalling firm
mo-Vis BVBA
Reason for recall
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Action taken
An URGENT: MEDICAL DEVICE RECALL R-NET JOYSTICKS notification dated 07/10/2025 was sent to consignees instructing them to upgrade affected devices according to the provided Annex 2 or by emailing contact@mo-vis.com to coordinate return. The software upgrade of affected devices should be conducted during the next scheduled annual maintenance of the joystick. These actions should be verified. Questions can be directed via email to contact@mo-vis.com or by phone at +32 9 335 28 60.
Root cause
Device Design
Status
Open, Classified
Product code
KNN
Quantity affected
3 units
Distribution
US Nationwide distribution in the state of TX.
Date initiated
2025-07-10
Date posted
2025-09-02
Location
Nevele, Belgium

Monitor recalls programmatically