Mucosal Atomization Device (MAD), Intranasal Mucosal Atomization Device with 3mL Syringe and Vial Adapter; Product Code
FDA device recall Z-0898-2017 · posted 2016-11-09 · Terminated
Recall details
- Recalling firm
- Teleflex Medical
- Reason for recall
- Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication.
- Action taken
- Teleflex sent an Urgent Medical Device Recall Notification letter on/about October 27, 2016. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately discontinue use and quarantine affected products. Customers were asked to complete the enclosed Recall Acknowledgement Form and fax it to Teleflex Customer Service or email to recalls@teleflex.com. For questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990. Teleflex sent an Amended Urgent Medical Device Recall Notification letter dated November 22, 2016, to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to continue to return affected product for a full refund per the procedure set forth in the letter of October 27, 2016. Some customers have indicated that, due to medical necessity, they plan to continue using the affected lots rather than return them. Teleflex is advising all such customers to follow the supplemental instructions for Use included as Attachment B. These supplemental instructions allow non-destructive testing of each unit prior to the procedure to determine if it is defective. Institutions that intend to continue using affected lots with this supplemental testing, please send the acknowledgment form to recall@teleflex.com with their name and contact information and the lots that they will continue to use, or call 866-246-6990. Customers with questions should contact their local sales representative or Customer Service at 1-866-246-6990.
- Root cause
- Process control
- Status
- Terminated
- Product code
- CCQ
- Quantity affected
- 2,674,650 units in total
- Distribution
- Worldwide distribution. U.S. (Nationwide) and Internationally to Canada, Belgium, Singapore and Thailand.
- Date initiated
- 2016-10-27
- Date posted
- 2016-11-09
- Date terminated
- 2020-09-25
- Location
- Research Triangle Park, NC