MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504),
FDA device recall Z-2041-2024 · posted 2024-06-09 · Open, Classified
Recall details
- Recalling firm
- Canon Medical System, USA, INC.
- Reason for recall
- The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
- Action taken
- On 5/10/24, correction notices were sent to customers who were asked to do the following: 1) The replacement of the power cable securing parts that may cause smoke and fire, with parts made of materials with a higher degree of flame retardance, is now being released. 2) A firm representative will contact you to schedule a time to replace the power cable securing parts. 3) If any abnormalities such as unusual smells or smoke are observed before the work, immediately stop using the system and contact your service representative. 4) The parts required to address the defect are expected to be available no earlier than June 2024. 5) Complete and return the reply form via DocuSign or via email to pbiggins@us.medical.canon For questions, please contact our InTouch team at intouchdispatch@us.medical.canon
- Root cause
- Device Design
- Status
- Open, Classified
- Product code
- LNH
- Quantity affected
- 153
- Distribution
- US: PA, SC, IA, AL, MI, AR, SD, TX, ID, KY, NY, MN, MD, OH, WI, NV, CT, AZ, IL, OK, CA, FL, GA, OR, IN, NJ, CO, MT, KS, MO, WY, VA, VT, LA, NH, WV, TN, PR
- Date initiated
- 2024-05-10
- Date posted
- 2024-06-09
- Location
- Tustin, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| Vantage Galan 3T, MRT-3020, V9.0 with AiCE Reconstruction Processing Unit for MR (K230355) | Canon Medical Systems Corporation | 2023-08-30 |
| Vantage Fortian/Orian 1.5T, MRT-1550, V8.0 with AiCE Reconstruction Processing Unit for MR (K222968) | Canon Medical Systems Corporation | 2022-10-25 |
| Vantage Elan 1.5T, MRT-2020, V6.0 (K210164) | Canon Medical Systems Corporation | 2021-03-10 |
| Vantage Titan 3T, MRT-3010, V2.5 (K191128) | Canon Medical Systems Corporation | 2019-06-26 |
| VANTAGE TITAN WITH HELIOS GRADIENT MODEL MRT-1504/US (K122613) | Toshiba America Medical Systems, In.C | 2012-12-13 |