Atlas FDA

Mouse anti-TTF-1 2nd Gen Predilute used in a panel designed to differentiate between different types of cancer when used

FDA device recall Z-0015-05 · posted 2004-10-09 · Terminated

Recall details

Recalling firm
Zymed Laboratories Inc.
Reason for recall
Mouse anti-TTF-1 had become contaminated by immunoglobulins from mouse serum that was co-eluted during purification leading to false readings.
Action taken
The firm has conducted telephone calls and visits to its consignees, with confirmation done by phone and fax. The firm has recovered or accounted for 100% of the products, either returned or destroyed.
Root cause
Other
Status
Terminated
Product code
MXZ
Quantity affected
24 vials of the mouse anti-TTF-1 concentrate have been released in distribution.
Distribution
The Firm distributed to a total of 8 domestic consignees and an unspecified number of foreign consignees. The products were distributed to MD, GA, TX, OH, IL, DC, NY and CA. Foreign countries distribu
Date initiated
2004-04-06
Date posted
2004-10-09
Date terminated
2004-10-29
Location
South San Francisco, CA