Atlas FDA

MOSAIQ, oncology information system - Product Usage: MOSAIQ is an oncology information system used to manage workflows f

FDA device recall Z-1597-2020 · posted 2020-03-26 · Open, Classified

Recall details

Recalling firm
Elekta Impac Software
Reason for recall
The user may inadvertently enter Metric values into Height and Weight fields labeled with US Standard Units.
Action taken
The firm began notifying their consignee on 09/24/2019 by email. The notice explained the issue and provided a workaround for the consignee to use until the updated software is ready for release.
Root cause
Software design
Status
Open, Classified
Product code
MUJ
Quantity affected
3732 units
Distribution
Worldwide distribution.
Date initiated
2019-09-24
Date posted
2020-03-26
Location
Sunnyvale, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
MOSAIQ ONCOLOGY INFORMATION SYSTEM (K141572)Impac Medical Systems, Inc.2014-09-05