MOSAIQ MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It support

FDA device recall Z-1869-2014 · posted 2014-06-25 · Terminated

Recall details

Recalling firm
Elekta, Inc.
Reason for recall
MOSAIQ does not re-calculate the agent volume when the user changes the ordering dose (to other than what was calculated) prior to saving the order.
Action taken
Elekta issued an Important Safety Notice dated May 29, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Corrective action that shall be taken: Notify the affected customers of the issue - To be completed 06/22/2014 Follow up with the affected customers for: Confirmation receipt of the LIN, AND Confirmation of installation of patch for the fix- To be completed 11/30/2014 The notice includes a confirmation which is to be returned. For further questions call (770) 670-2422.
Root cause
Software design
Status
Terminated
Product code
IYE
Quantity affected
101
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, IA, KY, LA, ME, MD, MA, MI, MO, MT, NH, NJ, NM, NY, NC, ND, OH, OK, OR
Date initiated
2014-05-29
Date posted
2014-06-25
Date terminated
2021-07-16
Location
Atlanta, GA