Atlas FDA

Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack, REF 1188100777

FDA device recall Z-0310-2025 · posted 2024-11-08 · Open, Classified

Recall details

Recalling firm
Cardinal Health 200, LLC
Reason for recall
Certain lots of listed lots of Cardinal Health Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack have demonstrated incompatibility with needleless Intravenous (IV) connectors. Cardinal Health recommends that they not be used for administration of IV push insulin via needleless connector.
Action taken
Cardinal Health issued an URGENT MEDICAL DEVICE PRODUCT RECALL to its consignees on 09/25/2024 via overnight mail. The notice explained the issue, risk to health, and requested the following actions: 1. REVIEW inventory for the affected product code and lots. 2. COMMUNICATE with all personnel who utilize the Cardinal Health Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack that it should not be used for administration of IV push insulin via needleless connector. 3. POST a copy of this notification in your storeroom where the product is stored. 4. NOTIFY any customers to whom you may have distributed/forwarded affected product or to whom you plan to send the product about this medical device product correction and share a copy of this notice. 5. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to GMB-FieldCorrectiveAction@cardinalhealth.com, whether you have affected product or not. CONTACT the appropriate Customer Service group with questions related to this notification. Monday Friday between 8:00am - 5pm EST: " Hospital 800-964-5227 " Federal Government 800-444-1166 " Distributor 800-635-6021 " All Other Customers 888-444-5440
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
MEG
Quantity affected
371,326 units
Distribution
Nationwide
Date initiated
2024-09-25
Date posted
2024-11-08
Location
Waukegan, IL