MoniTorr 10100 10-100 CSF Drainage System w/Patient Line One Way Valve MoniTorr 10102 10-102 CSF Drainage System w/Patie
FDA device recall Z-1310-2019 · posted 2019-05-20 · Terminated
Recall details
- Recalling firm
- Integra LifeSciences Corp.
- Reason for recall
- The firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount stopcock and MoniTorr Panel Mount stopcock.
- Action taken
- On April 8, 2019, the firm distributed Urgent Medical Device Correction letters to its affected customers. The communication advised customers of the possible breakage of the stopcock. Customers were advised to take the following precautions: When connecting a fluid-filled transducer (FFT) to the Panel Mount/Transducer Mount stopcocks: " Do not apply excessive force while attaching the FFT to the stopcock. " Take care during the priming process to not torque or bend the syringe and FFT. " Ensure luer lock is "finger tight;" do not over-tighten connection at the stopcock junction. When managing an external ventricular or lumbar drain (EVD) with a FFT attached: " Take care when manipulating the stopcock to zero the FFT and to obtain pressure readings, do not apply excessive force to the stopcock junction. " Use caution during patient transportation and during routine ICU care, to avoid direct contact of the FFT with other devices or equipment. " If the FFT must be flushed or replaced, do not apply excessive torque or force to the FFT/ stopcock junction. " If the FFT is connected to the ICU bedside monitor, ensure the interface cable is positioned so that it does not place excess force on the FFT (e.Q. do not allow cable to touch the floor. Customers were asked to determine if they had any affected product on hand, review and understand the precautions, and complete and return a completed acknowledgement form to the firm. Distributors were asked to forward the customer letter to any customers who received the affected product. Should you have any questions regarding these instructions, please contact Customer Service Monday to Friday 8:00 AM - 8:00 PM EST USA: 1-800-654-2873: email: custsvcnj@integralife.com Outside USA: 781-565-1401 ***UPDATE*** On October 22, 2019, the firm sent out an updated notification informing customers that a new stopcock has been designed for the LimiTorr and MoniTorr, and is now available for replacement of th
- Root cause
- Device Design
- Status
- Terminated
- Product code
- JXG
- Quantity affected
- 42102 units
- Distribution
- Worldwide Distribution - US Nationwide International distribution to Argentina, American Samoa, Australia, Canada, and Guam.
- Date initiated
- 2019-04-08
- Date posted
- 2019-05-20
- Date terminated
- 2021-01-27
- Location
- Plainsboro, NJ
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| MONITORR ICP EXTERNAL DRAINAGE AND MONITORING SYSTEMS (K062599) | Integra LifeSciences Corporation | 2006-09-27 |
| MONITORR ICP EXTERNAL CSF DRAINAGE AND MONITORING SYSTEM (K022554) | Integra Neurosciences | 2002-08-23 |
| CLINICAL NEURO SYST MONITORR EXT CSF DRAIN/MONI SY (K920156) | Clinical Neuro Systems, LLC | 1992-03-16 |