Monarch Inflation Device MAP 112 Access-9 Large Bore Hemostasis Value Metal Insertion Tool Torque DeviceREF IN2230 Injec
FDA device recall Z-0058-2025 · posted 2024-10-11 · Open, Classified
Recall details
- Recalling firm
- Merit Medical Systems, Inc.
- Reason for recall
- Due to a manufacturing process resulting in holes/breaches in the Tyvek sterile barrier.
- Action taken
- On 08/30/2024, the firm email and/or UPS overnight delivery an "URGENT MEDICAL DEVICE RECALL NOTICE" informing customers that specific lots of MONARCH Inflation Devices due to a packaging defect, there may be a small hole in the Tyvek sterile barrier of the device packaging. Use of the affected product may result in exposure to pathogens/pyrogens which could lead to infection. Customer are instructed to: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 4. Please fill out, scan, and email the completed CRF to Customer Service at response@merit.com within 7 calendar days. All affected product shipped to you must be accounted for on the CRF. 5. Please immediately return all affected lots in your possession to Merit, per the instructions in the attached CRF. For questions - contact your Merit Sales Representative or Merit Customer Service via email at RESPONSE@merit.com or via phone at +1 800 356 3748 | Hours: 6 am to 6 pm MST | Mon-Fri.
- Root cause
- Nonconforming Material/Component
- Status
- Open, Classified
- Product code
- DXT
- Quantity affected
- 41,411 devices
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR P
- Date initiated
- 2024-08-30
- Date posted
- 2024-10-11
- Location
- South Jordan, UT