Monaco Radiation Treatment Planning Workstation. Monaco Release 2.04.00 and above. Used to create treatment plans for an

FDA device recall Z-2569-2011 · posted 2011-06-15 · Terminated

Recall details

Recalling firm
Computerized Medical Systems Inc
Reason for recall
Monaco: In XiO IMRT (Dynamic MLC Delivery) software, the radiation beam angle is set to zero. If different beam angles are selected via Monaco software, the beam angles will remain at zero.
Action taken
Computerized Medical Systems, Inc. sent a User Notice to the affected consignees on October 27, 2010, explaining the reason for correction, the clinical impact, and the workaround method. (VERIFY USE NOTICE DATE) A return postcard was included for the customer to confirm receipt of the User Notice. The firm intends for this issue to be corrected in Monaco Release 3.00 in March 2011.
Root cause
Software design
Status
Terminated
Product code
MUJ
Quantity affected
11
Distribution
Worldwide Distribution - USA including CA, IL, MN, NM, OH, NY, TX, and WA and the Netherlands.
Date initiated
2010-10-27
Date posted
2011-06-15
Date terminated
2014-04-02
Location
Maryland Heights, MO