Atlas FDA

MODULAR CUP NEUTRAL LINER MODULAR CUP 10 DEGREE LIN MODULAR CUP 20 DEGREE LIN MODULAR CUP 7MM OFFSET LI

FDA device recall Z-1023-2014 · posted 2014-02-20 · Terminated

Recall details

Recalling firm
Zimmer, Inc.
Reason for recall
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
Action taken
Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.
Root cause
Packaging
Status
Terminated
Product code
LPH
Quantity affected
7,044,680 total
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, N
Date initiated
2013-10-11
Date posted
2014-02-20
Date terminated
2017-04-14
Location
Warsaw, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
NEXGEN COMPLETE KNEE SOLUTION MIS TIBIAL COMPONENTS, LOCKING SCREW AND STEM EXTENSIONS (K043101)Zimmer, Inc.2005-03-07
NEXGEN COMPLETE KNEE SOLUTION LEGACY POSTERIOR STABILIZED (LPS); LPS-FLEX FIXED BEARING FEMORAL AND ARTICULAR SURFACE CO (K991581)Zimmer, Inc.1999-07-30
NEXGEN COMPL. KNEE SOL. LEGACY POSTERIOR STABILIZED/CONSTRAINED CONDYLAR KNEE FEMORAL COMPONENTS/ARTICULAR SURFACES (K960279)Zimmer, Inc.1996-04-26
NEXT GENERATION KNEE SYSTEM (K933785)Zimmer, Inc.1995-01-30
INSALL/BURSTEIN* II MODULAR TOTAL KNEE (K872379)Zimmer, Inc.1987-10-28