Model #842-326 (pump tube for waste) & Model #842-327 (pump tube for solution), Waste Tubing used on ABL 7xx and ABL8xx,

FDA device recall Z-1260-06 · posted 2006-07-26 · Terminated

Recall details

Recalling firm
Radiometer America Inc
Reason for recall
An increased failure rate was observed with the yellow rubber tubing. The tubing was found to be leaking blood and solutions after distribution.
Action taken
The recalling firm issued a letter to their customers dated 5/22/2006. The letter informs the customer(s) of the tubing defect and provides the customer with one (1) tubing to deal with any immediate needs which they may have. Should the customer require additional waste pump tubing replacements, the letter is accompanied by a FAX FORM which can be used to request any amount of additional replacement tubing required. The letter also shows tthe address to which the replacement tubing is to be delivered, and provides a Purchase Order request to be used, if required.
Root cause
Other
Status
Terminated
Product code
CHL
Quantity affected
2500 Units in U.S. & 76 Units in Canada
Distribution
Nationwide to user/customers in the U.S. and in Canada
Date initiated
2006-05-22
Date posted
2006-07-26
Date terminated
2012-04-17
Location
Westlake, OH