Atlas FDA

Model 208-70-110, Long Thread Tiger Large Cannulated Headless Screw, 7 x 110 mm labeled as "7.0 x 110mm LTTHCS REF 208-7

FDA device recall Z-0609-2017 · posted 2016-11-17 · Terminated

Recall details

Recalling firm
Trilliant Surgical Ltd.
Reason for recall
Screws were identified as having the incorrect part number and lot number marked upon them. The part number incorrectly indicated a length of 115 mm rather than 110 mm. The lot number incorrectly indicated lot number of TSL003765 rather than TSL003721.
Action taken
The firm sent out written notifcation of the recall on 9/2/16; the letter instructed the consignees to discontinue use of the affected lot, quarantine, and return product of the affected lot in accordance with an included acknowledgement sheet.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
HWC
Quantity affected
11
Distribution
CA, AZ, WI, FL, NJ, OH
Date initiated
2016-09-02
Date posted
2016-11-17
Date terminated
2017-05-19
Location
Houston, TX

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
TIGER HEADLESS CANNULATED SCREWS (K112737)Trilliant Surgical, Ltd.2011-10-20