Mobile Column TruSystem 7500, Material Number 1717023; Mobile Column TruSystem 5500, Material Number 1607822 Patient pos
FDA device recall Z-0040-2017 · posted 2016-10-05 · Terminated
Recall details
- Recalling firm
- Trumpf Medical Systems, Inc.
- Reason for recall
- Three complaints describing TruSystem tables unexpectedly descending approximately 20 centimeters.
- Action taken
- Trumpf Medical sent an Urgent Field Safety Notice dated July 8, 2015, to all affected consignees. The letter indicated that the tables should not be used until an authorized service engineer has exchanged the spindles of the affected operating tables. The letter included an acknowledgement form which is to be returned. The letter also requested a sub-recall if the product had been further distributed. For questions regarding this recall call 843-329-0543.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- JEA
- Quantity affected
- 84 units
- Distribution
- Worldwide Distribution - US including CA, CT, GA, IL, NC, OR, SC, VA, WI, WY, and Internationally to Australia, Azerbaijan, Brazil, Finland, France, Germany, Hungary, Italy, Luxembourg, Norway, Poland
- Date initiated
- 2016-07-10
- Date posted
- 2016-10-05
- Date terminated
- 2017-03-02
- Location
- Charleston, SC