Misys Laboratory System Version 5.3
FDA device recall Z-0670-03 · posted 2003-03-21 · Terminated
Recall details
- Recalling firm
- Misys Healthcare Systems
- Reason for recall
- Software anomaly. The HIDE command used to suppress suspected individual microbiology observations does not work in GUI inquiry
- Action taken
- Notification was made by FAX to customers. A procedureal work around was provided also. The firm will announce the availability of a software fix in the future.
- Root cause
- Other
- Status
- Terminated
- Product code
- JQP
- Quantity affected
- 464
- Distribution
- Nationwide and to United Kingdom, Ireland, Canada, Denmark, Bermuda, Saudi Arabia, United Arab Emirates.
- Date initiated
- 2003-02-19
- Date posted
- 2003-03-21
- Date terminated
- 2004-06-02
- Location
- Tucson, AZ