Minimed Model 503 Remote Transmitter (MMT-503, MMT-503EU, MMT-503NA, MMT-503US) for use with the MiniMed Paradigm Insuli
FDA device recall Z-0178-2020 · posted 2019-11-04 · Terminated
Recall details
- Recalling firm
- Medtronic Inc.
- Reason for recall
- There is a potential security vulnerability related to the use of the remote controller accessories with the insulin pumps.
- Action taken
- Medtronic sent an Urgent Field Safety Notification letter dated August 7, 2018 to the patient, distributor, and healthcare provider via email, U.S. mail, or phone. The letter identified the affected product, problem and actions to be taken. Starting on 10/05/2021, recall notices were sent to patients, health care professionals, distributors, and personnel at clinical sites. Distributors were asked to provide their customer lists to the recalling firm so the recalling firm could send customers recall notices. The scope of the recall is being expanded to all users regardless of warranty status. Users are asked to stop using and disconnect the remote controller, disable the remote feature, and return the remote controller to the recalling firm. Customers were also asked to provide the information requested in the confirmation form to the recalling firm via phone, mail, email, or online on the firm's website. Additional questions can be directed to Technical Support at 1-800-378-2292. Disconnecting/Disabling instructions were provided in the recall notice, which provided the following link: https://info.medtronicdiabetes.com/remote-controller-return
- Root cause
- Software design
- Status
- Terminated
- Product code
- LZG
- Quantity affected
- 141,519
- Distribution
- Distribution was nationwide to consumers and distributors. There was no government/military distribution. Foreign distribution was made to Algeria, Australia, Austria, Belgium, Brazil, Canada, China,
- Date initiated
- 2018-08-07
- Date posted
- 2019-11-04
- Date terminated
- 2024-03-29
- Location
- Northridge, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| MEDTRONIC MINIMED PARADIGM INSULIN PUMPS (K120206) | Medtronic Minimed | 2012-03-29 |
| MINIMED PARADIGM INSULIN PUMP, MODEL 511 (K001829) | Minimed, Inc. | 2000-08-18 |