Mini MaxLock Extreme Plating System Case & Tray, Model Number(s): MXM-901, Product Usage: The Mini MaxLock Extreme Plati
FDA device recall Z-3081-2011 · posted 2011-08-29 · Terminated
Recall details
- Recalling firm
- Orthohelix Surgical Designs Inc
- Reason for recall
- The surgical tray does not meet our Instruction for Use (IFU) for dry time requirement.
- Action taken
- Orthohelix Surgical Designs Inc sent a RECALL NOTIFICATION letter dated July 6, 2011 and an UPDATED URGENT RECALL 2ND NOTIFICATION letter dated July 13, 2011 to all affected customers. The letter identified the recalled product, the problem, and the Field Correction actions to be taken. Customers were instructed to remove the blue rubber mat in the utility bay of the tray base immediately, tag the mat with the tray number, and return it to recalling firm immediately. The letter states once customers have removed the blue rubber mat from the tray base, they may continue to use the tray and system normally. No further actions or precautions are needed. Distributors were also asked to notify their customers of the recall. If you have any questions, please call 330-247-1444 or 330-247-1445.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- HRS
- Quantity affected
- 61 trays
- Distribution
- USA Nationwide Distribution including the states of: AR, AZ, CA, CO, FL, GA, IL, IN, KY, MA, MD, MI, MO, NC, NE, NJ, NM, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WI & WY.
- Date initiated
- 2011-07-06
- Date posted
- 2011-08-29
- Date terminated
- 2012-09-10
- Location
- Medina, OH