Atlas FDA

MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidif

FDA device recall Z-1377-2018 · posted 2018-04-13 · Terminated

Recall details

Recalling firm
Cook Inc.
Reason for recall
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
Action taken
Cook plans to begin notifying consignees of the recall on January 12, 2018 Written recall communication letters will be mailed via courier to US consignees
Root cause
Device Design
Status
Terminated
Product code
MQG
Quantity affected
483
Distribution
Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)
Date initiated
2018-01-12
Date posted
2018-04-13
Date terminated
2020-05-08
Location
Bloomington, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
COOK MINI-INCUBATOR (K983642)Cook Ob/Gyn1999-02-01