Miltex Premium Grade Self-Lock Chisel Blade Handle, Catalogue No. 4-405, manual surgical instrument.
FDA device recall Z-1841-2016 · posted 2016-05-31 · Terminated
Recall details
- Recalling firm
- Integra York PA, Inc
- Reason for recall
- One lot of self-lock chisel blade handles have a wider slot that holds the chisel blade that may result in the inability to firmly secure the blade in the handle.
- Action taken
- Integra mailed out an Urgent: Voluntary Medical Device Recall letter to customers informing them to evaluate their inventory for the affected product. If product was shipped to customers, they were asked to identify their customers and to stop shipping immediately. Integra requested that customers complete the attached form that was sent with the notification indicating whether or not they had the affected product on hand and customer service will contact them to provide a Return Authorization number for the product's return. Customers could reach Customer Service at 866-854-8300.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- GDZ
- Quantity affected
- 678 Chisel Blade Holders
- Distribution
- Worldwide distribution. US nationwide, Mexico, and Korea.
- Date initiated
- 2016-03-11
- Date posted
- 2016-05-31
- Date terminated
- 2017-04-17
- Location
- York, PA