Atlas FDA

Millenium VG with Hawkeye/Discovery VH with Hawkeye, model number: BR321000008. GE Healthcare, Waukesha, Wisconsin 53118

FDA device recall Z-0496-2010 · posted 2010-01-21 · Terminated

Recall details

Recalling firm
GE Medical Systems, LLC
Reason for recall
Failure to provide the complete User Manual information.
Action taken
A Field Modification Instruction will be released in February 2010. GE Healthcare will correct the Operator Manuals in forward production and the installed base units.
Root cause
Radiation Control for Health and Safety Act
Status
Terminated
Product code
JAK
Quantity affected
343 total
Distribution
Nationwide Distribution
Date initiated
2009-08-01
Date posted
2010-01-21
Date terminated
2012-02-04
Location
Waukesha, WI