MilCAM Recon III Model Family of IR Imagers that incorporate a Class I infared laser for Laser Range Finding (LRF).
FDA device recall Z-0937-2009 · posted 2009-09-25 · Terminated
Recall details
- Recalling firm
- Flir Systems, Inc.
- Reason for recall
- LRF models were sold to non-DoD customers prior to certification and reporting as required in Title 21 of the Code of Federal Regulations, Subchapter J, Parts 1000 - 1050 (21 CFR 1000 - 1050). A filter defect allows transmission of laser energy such that the affected LRF would be Class 3B laser product, however the product is reported to CDRH as a Class I product.
- Action taken
- Customers will be notified by a letter instructing them to stop using the LRF, contact a Customer Service Representative, and to detach and return the Laser Range Finder. For each returned LRF, FLIR will provide a certified Class I LRF.
- Root cause
- Radiation Control for Health and Safety Act
- Status
- Terminated
- Product code
- REQ
- Quantity affected
- 120
- Distribution
- Nationwide Distribution
- Date initiated
- 2009-01-30
- Date posted
- 2009-09-25
- Date terminated
- 2015-06-02
- Location
- North Billerica, MA