Microstream SMART CAPNOLINE PLUS ADU/INT 25UN Part Number: 009818 Oridion 3012176-011 Physio Control

FDA device recall Z-2003-2010 · posted 2010-07-13 · Terminated

Recall details

Recalling firm
Oridion Medical 1987 Ltd.
Reason for recall
The Microstream Filterline may not be recognized by the monitor when connected. No CO2 readings will be displayed and no alarms may sound for unmonitored patients. Replacing the filterline may delay treatment.
Action taken
Oridion notified customers by letter on 5/26/10 and requested that they remove the FilterLines from the related production lots from the field and an immediate replacement with FilterLines from subsequent production lots. Questions are directed to the responsible person at the supplier named in the letter, or to Oridion Capnography at 888-674-3466.
Root cause
Process design
Status
Terminated
Product code
cck
Quantity affected
2961 boxes
Distribution
AZ, CA, FL, HI, LA, , MA, MO, NJ, OH, PA, WA, WI,
Date initiated
2010-05-26
Date posted
2010-07-13
Date terminated
2012-06-21
Location
Jerusalem, Har Hotzvim Industrial Park, Israel