MicroScan WalkAway-40 plus Instrument and MicroScan Walkaway-96 plus instrument access door hinge. Siemens Healthcare Di

FDA device recall Z-1990-2014 · posted 2014-07-02 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
Springs contained in the access door hinge assembly on the Walk Away Plus instruments are becoming detached causing premature failure of the hinge assembly and a potential safety issue for the customers.
Action taken
Urgent Medical Device Recall notification letter was sent to all affected US customers on June 2nd by Federal Express. OUS notifications were also sent to regional Siemens organizations who distributed in-country notifications and relevant notifications.
Root cause
Component design/selection
Status
Terminated
Product code
LRG
Quantity affected
WA40: 560 US, 131 OUS, 9 already repaired; WA96: 81 US, 189 OUS, 57 repaired.
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Algeria Argentina Brazil Canada China Chile Columbia France Indonesia Libya Georgia Germany India Italy Japan Kazakhstan Lithuania Oman Pak
Date initiated
2014-06-02
Date posted
2014-07-02
Date terminated
2015-12-23
Location
West Sacramento, CA